On the TPS2000+ where is the trigger point when the pulse width trigger is used and a specified amount of time is selected?

Question :

On the TPS2000+ where is the trigger point when the pulse width trigger is used and a specified amount of time is selected?

Answer :

Equal: Oscilloscope triggers when trailing edge of the pulse crosses the trigger level.

Not equal: If the pulse is narrower than the specified width, the trigger point is the trailing edge. Otherwise, the oscilloscope triggers when a pulse continues longer than the time specified as the pulse width.

Less than: The trigger point is the trailing edge.

Greater than (also called time-out trigger): The oscilloscope triggers when a pulse continues longer than the time specified as the pulse width.